Precision Reference Thermometer
|
Industrial Professional Engineering Services
|
Precision Reference Thermometer
|
- Adjustable Pressure Relief Valve PRV-12 for direct mounting on Hand Held Test Pump HTP1.
- Protects Reference and Unit under Test.
The PRV-12 range of relief valves
provides a compact solution to the
protection of sensitive instruments
from the danger of overpressure
under low flow conditions.
The stainless steel body houses a
piston which is retained by a series
of disc springs, the number and
configuration of which control the
'cracking' pressure required (see
table below).
The PRV-12 is for use only with the HTP1 Hydraulic Pressure Test System. The
PRV-12 is simply screwed into the rear 1/4" BSP blanked off port which is reserved
for the purpose. The valve can be supplied pre-set with the HTP1 or can be quickly
retro-fitted to existing units using simple installation instructions.
Specification:
Material: Stainless steel AISI 303
Seals: Polyurethane, PTFE, Nitrile, Nylon
Fluid media: Mineral Oill, destilled water, optional "Skydrol", other media on request
Max. working pressure: 1000 bar
Dimensions: 40 mm x 22 mm Diameter
Weight: appr. 60 g
Accuracy / Repeatability: +/- 10% nom. setting
Type 'Cracking' Pressure Range Order-Code
PRV-12-1 10 to 50 bar 210.9.000.0700.6
PRV-12-2 50 to 200 bar 210.9.000.0700.7
PRV-12-3 200 to 400 bar 210.9.000.0700.8
PRV-12-4 300 to 700 bar 210.9.000.0700.9
PRV-12-5 600 to 1000 bar 210.9.000.0701.0
Pressure adjustment is achieved by first removing the protective cap, then using a
suitable tool, turn the adjustment screw clockwise to increase and anticlockwise to
decrease 'cracking' pressure.